Large range and high resolution on-line
displacement measurement system by combining
double interferometres
Fang Xie,* Min Li, Ding Song, Jinling Sun, and Tao Zhang
Laboratory for Optical Information Science and Technology, Department of Physics, School of Science,
Beijing Jiaotong University, Beijing 100044, China
*fxie@bjtu.edu.cn
Abstract: A stabilized interferometric displacement measurement system,
which is suitable for on-line measurement and is endowed with large
measurement range and high resolution, is proposed. The system is
stabilized by a feedback loop which compensates the influences induced by
the environmental disturbances and makes the system stabile enough for on-
line measurement. Two different wavelengths are working simultaneously
in the system. The measurement range which is determined by the synthetic-
wavelength interferometric signal is expanded to the order of millimeter,
while the measurement resolution which is determined by one of the single-
wavelength interferometric signal is the order of sub-nanometer.
©2010 Optical Society of America
OCIS codes: (060.2370) Fiber optics sensors; (120.3180) Interferometry; (120.0120)
Instrumentation, measurement, and metrology.
References and links
1. D. P. Hand, T. A. Carolan, J. S. Barton, and J. D. C. Jones, “Profile measurement of optically rough surfaces by
fiber-optic interferometry,” Opt. Lett. 18(16), 1361–1363 (1993).
2. D. A. Jackson, R. Priest, A. Dandridge, and A. B. Tveten, “Elimination of drift in a single-mode optical fiber
interferometer using a piezoelectrically stretched coiled fiber,” Appl. Opt. 19(17), 2926–2929 (1980).
3. D. Lin, X. Jiang, F. Xie, W. Zhang, L. Zhang, and I. Bennion, “High stability multiplexed fiber interferometer
and its application on absolute displacement measurement and on-line surface metrology,” Opt. Express 12(23),
5729–5734 (2004).
4. M. Jiang, and E. Gerhard, “A simple strain sensor using a thin film as a low-finesse fiber-optic Fabry-Perot
interferometer,” Sens. Actuators A Phys. 88(1), 41–46 (2001).
5. H. C. Seat, E. Ouisse, E. Morteau, and V. Metivier, “Vibration-displacement measurements based on a
polarimetric extrinsic fiber Fabry-Perot interferometer,” Meas. Sci. Technol. 14(6), 710–716 (2003).
6. B. T. Meggitt, C. J. Hall, and K. Weir, “An all fibre white light interferometric strain measurement system,”
Sens. Actuators A Phys. 79(1), 1–7 (2000).
7. L. V. Nguyen, D. Hwang, S. Moon, D. S. Moon, and Y. Chung, “High temperature fiber sensor with high
sensitivity based on core diameter mismatch,” Opt. Express 16(15), 11369–11375 (2008).
8. W. Wang, N. Wu, Y. Tian, C. Niezrecki, and X. Wang, “Miniature all-silica optical fiber pressure sensor with an
ultrathin uniform diaphragm,” Opt. Express 18(9), 9006–9014 (2010).
1. Introduction
Optical fiber interferometers have been widely used in metrology due to their prominent
advantages such as non-contact, compactness, light weight, immunity to electromagnetic
interference, high resolution, low cost. Hence there is a great amount of interest to exploit
optical fiber interferometers for the measurement of a large variety of parameters such as
displacement, vibration, velocity, strain, pressure and temperature [1–8].
For a fiber interferometric displacement measurement system, the fiber should be used just
for transmitting the light and the phase change in the interferometric signal should be just
induced by the measured displacement. However, the length of the fiber configuring the
interferometric arms will change randomly because of temperature fluctuations and other
types of the environmental disturbances, thus will induce low frequency random phase drifts
in the interferometric signal, which will decrease the measurement precision, and in the worst
#134128 - $15.00 USD Received 27 Aug 2010; revised 31 Oct 2010; accepted 31 Oct 2010; published 15 Nov 2010
(C) 2010 OSA 22 November 2010 / Vol. 18, No. 24 / OPTICS EXPRESS 24961
circumstances, will even make the measurement system unfunctioning. Although techniques
such as common-path interferometric arms [1] and feedback compensating methods [2,3] are
able to compensate the random phase drifts in interferometric signals, however, because of the
phase ambiguity in interferometric signal, the measurement range of [1] which is limited to a
half-wavelength is only several hundreds of nanometers. And this limitation makes the
systems unable to be used in many fields.
We propose a highly stabilized fiber Michelson interferometric displacement measurement
system which is endowed with large measurement range and high resolution. A feedback loop
compensates the random phase drifts in the interferometric signal which is induced by the
environmental disturbances and stabilizes the measurement system for on-line measurement.
The light from two temperature-stabilized DFB laser sources with different wavelengths are
coupled simultaneously into the interferometric measurement system. The two wavelengths
are guided into proper optical routes by employing a fiber Bragg grating(FBG)as an in-fiber
reflective mirror. A photo detector detects the single-wavelength interferometric signal while
another photo detector detects the synthetic-wavelength interferometric signal at the same
time. By combining and processing the two signals properly, the measurement range of the
system which is determined by the synthetic-wavelength interferometric signal can be
expanded to the order of millimeter, while the resolution which is determined by the single-
wavelength interferometric signal can be the order of sub-nanometer.
2. The principle of the system
2.1 The proposed fiber interferometer measurement system
The proposed fiber interferometer measurement system is shown in Fig. 1. Two temperature-
stabilized DFB laser sources with outputs at
1 1558.52 nm and 2 1557.32 nm are used in
the system. The lasers give an output power about 2.5mW with 20dB spectral bandwidth
0.2nm. The light emitted from the two lasers passes through 3dB-coupler 1, circulator 1, 3dB-
coupler 2 and is divided into two beams. The two beams are collimated by two grin lenses
respectively and are reflected back into the system by measurement mirror and reference
mirror. The two reflected beams, which are configurated with two different wavelengths, are
combinated again at 3dB-coupler 2 and interfere with each other and then the synthetic-
wavelength interferometric signal is produced. The synthetic-wavelength interferometric
signal from one port of 3dB-coupler 2 is guided by circulator 1 and is detected by photo
detector 1 (PD1). The interferometric signal from the other port of 3dB-coupler 2 is guided by
circulator 2 and reaches a fiber Bragg grating (FBG). As the Bragg wavelength of FBG is
chosen to be 1557.32 nm with 3dB bandwidth 0.3nm, the light with the wavelength of
1557.32nm will reflected by the FBG while the light with the wavelength of 1558.52nm will
transmit FBG and is guided out of the system. The light reflected by FBG, which is a single-
wavelength interferometric signal with the wavelength of 1557.32nm, is guided by circulator
2 again and is detected by photo detector 2 (PD2). The signals detected by PD1 and PD2 are
processed by the electronic processor simultaneously, and the signal detected by PD2 is input
into an electronic feedback loop at the same time. An information derived from this electronic
feedback loop is used as a correction signal and is applied on a stacked piezoelectric
transducer (PZT) on which the reference mirror is mounted. The feedback signal drives PZT
to tune the optical path of the reference arm in order to keep the interferometer in quadrature
state (the phase difference is 2/ ). The random phase drift with low frequency in the fiber
interferometer, which is induced by the environmental disturbances, is effectively eliminated
and therefore the fiber interferometer measurement system is stabilized. By the means of this,
the environmental disturbances will not influence the measurement system anymore.
When the measured displacement varies, the signals detected by PD1 and PD2 will vary
correspondently. The period of the signal detected by PD2 is half the wavelength
2 , while
the period of the signal detected by PD1 is half the synthetic wavelength of
1 and 2 , that is
#134128 - $15.00 USD Received 27 Aug 2010; revised 31 Oct 2010; accepted 31 Oct 2010; published 15 Nov 2010
(C) 2010 OSA 22 November 2010 / Vol. 18, No. 24 / OPTICS EXPRESS 24962
Fig. 1. The principle of the measurement system.
1 2 2 12( ) , which is much bigger than 1 2 and 2 2 . By employing these characteristics,
the present system uses the signal from PD1 to determine the measurement range and the
signal from PD2 to determine the measurement resolution. So the system is capable of
maintaining the resolution as high as the order of sub-nanometer while expanding the
measurement range to the order of millimeter.
2.2 The principle of the electronic feedback loop
During the experiments, the optical power emitted from the two DFB lasers has been adjusted
to be the same, both are 2.5mW. The signals detected by PD1 and PD2 can be expressed as
Eq. (1) and Eq. (2) respectively.
1 0 1 1 0 2 21 cos 1 cos ,PD d S d Si i k i k (1)
2 0 2 21 cos ,PD d Si i k (2)
where
0i is the optical-current which is related to the input optical power, k is the
interferometric fringe visibility,
1d and 2d are the static differential phase between the two
arms for the two wavelengths respectively, and 1s and 2s are the differential phase induced
by environmental disturbances for the two wavelengths respectively.
After processed by current-to-voltage converters respectively, the signals shown in Eq. (1)
and Eq. (2) have the forms that are shown in Eq. (3) and Eq. (4).
1 1 0 1 1 0 2 21 cos 1 cos ,PD d S d Su u k u k (3)
1 2 0 2 21 cos ,PD d Su u k (4)
where
0u is related to the gain of the current-to-voltage converters.
The interferometric signal from PD2 is also used to eliminate the random phase drifts with
an electronic feedback loop that is resulted from the environmental disturbances. The
schematic diagram of the circuit of electronic feedback loop is shown in Fig. 2. PD2 is
connected to current-to-voltage converter U1. After passing through the electronic
FBG
2
1
3dB-coupler
2 1
Grin lens
Circulator
Ref. mirror
Output
PD2 PD1
1
2
Electronic
processor
DFB1
PZT
Meas. mirror
A/D converter
DFB2
Signal
generator
M
Electronic
feedback
loop
#134128 - $15.00 USD Received 27 Aug 2010; revised 31 Oct 2010; accepted 31 Oct 2010; published 15 Nov 2010
(C) 2010 OSA 22 November 2010 / Vol. 18, No. 24 / OPTICS EXPRESS 24963
differentiator U2, the DC part in (1)
2PDu will be erased, and the output from U2 will be shown as
Eq. (5).
2 2 2 2sin ,PD d Su K (5)
where K is the conversion gain of the interferometric signal from U2. After passing through
the electronic integrator U3, the output from U3 is the value shown in Eq. (6).
3 2 1 2 2cos ,PD d Su K (6)
where
1K is the conversion gain of U3. In quadrature position, there will be 2 2
2
d s
, it
obtains (3)
2 0PDu and the feedback loop voltage applied on PZT is zero. In the vicinity of
quadrature position, Eq. (6) can be rewritten as Eq. (7).
4
2 2 ,PDu K (7)
where
2K is the conversion gain of the interferometric signal from U3 in the vicinity of
quadrature position, and
2 2( )
2
d s
.
(4)
2PDu is linear to the value the interferometer
excursion to quadrature state and is used as a correction signal applied on PZT. As soon as the
interferometer drifts away from the quadratrue state because of the influences from the
environmental disturbances, the correction signal will drive PZT to tune the optical path in the
reference arm and the interferometer will be drawn back to the quadratrue state again. The
interferometer is kept at the quadrature state and the measurement system is stabilized. The
feedback loop is a first order system with a frequency bandwidth ranging from 0 to 21.65Hz
which is verified by the experiments sufficient to eliminate the environmental disturbances.
+
C1
10pF
R2 200K
R1
100K
-
2
3
+
R4
1M
-
2
3
+
R7
100K
-
2
3
C3
1uF
PD2
R3
100K 1uF
C2
1M
R5
R6
100K
To PZT
U1
GND
U2
LF356
U3
LF356
LF356
Fig. 2. The schematic diagram of the feedback loop circuit.
In the qudrature state, the signals from PD1 and PD2 have the following forms that are
shown in Eq. (8) and Eq. (9) respectively.
2
1
2
0 0
1
1 cos ,
2
PD
u u k u
(8)
5
2 0.PDu u (9)
It can be known from Eq. (8) and Eq. (9) that both the signals detected by PD1 and PD2
are stabilized at constant values when the interferometer is locked in quadrature state by the
signal from PD2.
#134128 - $15.00 USD Received 27 Aug 2010; revised 31 Oct 2010; accepted 31 Oct 2010; published 15 Nov 2010
(C) 2010 OSA 22 November 2010 / Vol. 18, No. 24 / OPTICS EXPRESS 24964
3. The experimental results
3.1 The validation of the feedback control loop
With the feedback loop out of operation and with measurement mirror and reference mirror in
static state, the interferometric signals detected by PD1 and PD2 are shown in Fig. 3. Curve 1
is the signal detected by PD1 which is synthetic-wavelength interferometric signal, while
curve 2 is the signal detected by PD2 which is single-wavelength interferometric signal. It can
be seen that the interferometric signals are fluctuating at all the time even though
measurement mirror and reference mirror are not moving. The variation in the interferometric
signals are resulting from temperature fluctuations and other types of the environmental
disturbances. However, as soon as the feedback loop is turned on, the interferometric signals
detected by PD1 and PD2 are stabilized at constant values, just as shown in Fig. 4. In Fig. 4,
the feedback loop is out of operation during the first part of the recorded time, it can be seen
that the interferometric signals detected by PD1 and PD2 are fluctuating during this period.
And during the second part of the recorded time in Fig. 4, the feedback loop is turned on and
the signals detected by PD1 and PD2 are stabilized respectively at constant values promptly.
These results are confirmed precisely to the theoretical analysis that are shown in Eq. (8) and
Eq. (9). We had experimented continuously to turn on the feedback loop for more than 4
hours, it was shown that the interferometer had been stabilized robustly during the
experiments.
3.2 The measurement results
When there is displacement L in the position of measurement mirror, the signals from PD1
and PD2 are shown in Eq. (10) and Eq. (11).
3
1
2
0 0
1 1 2
2 2
1 cos 2 1 cos 2 ,
2 2
PD
L L
u u k u k
(10)
6
2 0
2
2
1 cos 2 .
2
PD
L
u u k
(11)
In order to realize the displacement measurement, the interferometric signals have been
modulated linearly. A ramp voltage with low frequency is applied on an one-dimension
translation stage M (PI product, P-611.1S) to tune linearly the optical path difference in the
measurement interferometer. The displacement of M is amplified 8 times linearly by a level
made of aluminum alloy to meet the needed scanning distance. The interferometric signals
detected by PD1 and PD2 will vary correspondently, which have the forms that are shown in
Eq. (12) and Eq. (13) respectively.
4
1
2
0 0
1 1 2
2 2
1 cos 2 1 cos 2 ,
2 2
PD
L ct L ct
u u k u k
(12)
7
2 0
2
2
1 cos 2 ,
2
PD
L ct
u u k
(13)
where c is the scanning velocity of M and t is the scanning time.
Equation (12) can be rewritten as Eq. (14).
5 1
2 1 2 1 2 1 2 1
0 0
1 1 2 1 1 2
1 1
2 cos 4 cos 4 .
2 2 2 2
PD
u u u k L ct L ct
(14)
#134128 - $15.00 USD Received 27 Aug 2010; revised 31 Oct 2010; accepted 31 Oct 2010; published 15 Nov 2010
(C) 2010 OSA 22 November 2010 / Vol. 18, No. 24 / OPTICS EXPRESS 24965
It is known from Eq. (14) that the amplitude
0u k is modulated by the factor
2 1 2 1
1 1 2
1
cos 4
2 2
L ct
.The amplitude will be the maximum when,
2 1 2 1
1 1 2
1
4 ,
2 2
L ct
(15)
where m is an integral.
The scanning distance that the peak of the signal from PD1 occurring is shown in Eq. (16).
1 2 2
0
2 1
.
2( ) 8
ct L
(16)
It can be seen from Eq. (17) that peak position is shifting linearly with the value of L
rigorously.
The variation in the signals from PD1 and PD2 during a scanning period of M is shown in
Fig. 5. Curve 1 is the synthetic-wavelength interferometric signal detected by PD1 while
curve 2 is the single-wavelength interferometric signal detected by PD2. As the frequencies of
the modulated interferometric signals from PD1 and PD2 are much higher than 21.65Hz, the
compensating action of the feedback loop will not influence the interferometric signals. The
amplitude of the ramp voltage is tuned to correspond to one period change in the synthetic-
wavelength interferometric signal. When the displacement of the measurement mirror
changes, the peak of curve 1 will shift linearly. And it is known from Eq. (13) that the phase
change of curve 2 between the beginning of scanning action and the position of the peak on
curve 1 is linear precisely with the displacement of the measurement mirror. In order to
measure the phase change of curve 2 and then calculate the displacement of the measurement
mirror, curve 1 and curve 2 are converted simultaneously into digital signals by an A/D
converter and the data is processed by a software in a PC.
Fig. 3. The interferometric signals are fluctuating when the feedback loop is out of operation.
Fig. 4. The interferometric signals before and after the feedback loop is turned on.
The position of the peak on curve 1 can be determined easily by the software, and the
correspondent phase change of curve 2 between the beginning of the scanning action and the
position of the peak on curve 1 can be also measured by software. The software calculates the
#134128 - $15.00 USD Received 27 Aug 2010; revised 31 Oct 2010; accepted 31 Oct 2010; published 15 Nov 2010
(C) 2010 OSA 22 November 2010 / Vol. 18, No. 24 / OPTICS EXPRESS 24966
amplitude of the signle-wavelength interferometric signal first, and decides sequentially each
data’s phase based on its value and finally calculates the total phase change during the
addressed period. The displacement can therefore be calculated. Figure 6 is the zoom of Fig. 5
in the area of the top of synthetic-wavelength interferometric signal. The measurement range
of the system is determined by the synthetic-wavelength interferometric signal which is a half
synthetic-wavelength. Determined by the specific wavelengths used in the system, the
measurement range of the system is calculated to be about 1mm. The used A/D converter is
16 bits. The amplitude of the input voltage of A/D converter is 10V and so its voltage
resolution