Designation: D 3233 – 93 (Reapproved 1998) An American National Standard
Standard Test Methods for
Measurement of Extreme Pressure Properties of Fluid
Lubricants (Falex Pin and Vee Block Methods)1
This standard is issued under the fixed designation D 3233; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 These test methods cover two procedures for making a
preliminary evaluation of the load-carrying properties of fluid
lubricants by means of the Falex Pin and Vee Block Test
Machine.
NOTE 1—Additional information can be found in Appendix X1 regard-
ing coefficient of friction, load gage conversions, and load gage cablibra-
tion curve.
1.2 The values stated in SI units are to be regarded as
standard. Because the equipment used in these test methods is
available only in inch-pound units, the SI units are omitted
when referring to the equipment and the test specimens.
1.3 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appro-
priate safety and health practices and determine the applica-
bility of regulatory limitations prior to use.
2. Referenced Documents
2.1 ASTM Standards:
B 16 Specification for Free-Cutting Brass, Rod, Bar, and
Shapes for Use in Screw Machines2
D 2670 Test Method for Measuring Wear Properties of
Fluid Lubricants (Falex Pen and Vee Block Method)3
D 2783 Test Method for Measurement of Extreme-Pressure
Properties of Lubricating Fluids (Four-Ball Method)3
3. Terminology
3.1 Definitions of Terms Specific to This Standard:
3.1.1 actual gage load, n—the value obtained from the gage
while running the test and before any corrections are made.
3.1.1.1 Discussion—This gage reading is irrespective of the
particular gage used, and corrections are made by comparison
to a standard reference.
3.1.2 direct load, n—that which is applied linearly, bisect-
ing the angle of the vee block corrected to either the 800 or
3000-lbf gage reference.
3.1.2.1 Discussion—This load is equivalent to the true load
times the cos 42°.
3.1.3 true load, n—the sum of the applied forces normal to
the tangents of contact between the faces of one vee block and
the journal pin corrected to the 4500 lbf gage reference line.
3.1.4 true load failure value, n—the true load at which the
lubricant tested can no longer support the applied load resulting
in either test pin or shear pin breakage, or inability to maintain
or increase load.
3.1.4.1 Discussion—This value is also referred to as the
limit of extreme pressure.
4. Summary of Test Methods
4.1 Both test methods consist of running a rotating steel
journal at 290 6 10 rpm against two stationary V-blocks
immersed in the lubricant sample. Load is applied to the
V-blocks by a ratchet mechanism. In Test Method A (Note 1),
increasing load is applied continuously. In Test Method B
(Note 1), load is applied in 250-lbf (1112-N) increments with
load maintained constant for 1 min at each load increment. In
both methods the load-fail value obtained is the criteria for
level of load-carrying properties. Both methods require cali-
bration of the load gage and reporting of test results as true
(corrected) loads rather than actual gage loads.
NOTE 2—Test Method A is referred to as the Falex Run-Up Test. Test
Method B is referred to as the Falex One-Minute Step Test.
5. Significance and Use
5.1 Evaluations by both test methods differentiate between
fluids having low, medium, and high levels of extreme-pressure
properties. The user should establish any correlation between
results by either method and service performance.
NOTE 3—Relative ratings by both test methods on the reference fluids
covered in Table 1 and Table 2 are in good general agreement with
four-ball weld-point relative ratings obtained on these same reference
fluids, covered in Test Method D 2783.
1 These test methods are under the jurisdiction of ASTM Committee D-2 on
Petroleum Products and Lubricants and are the direct responsibility of Subcommit-
tee D02.0L on Industrial Lubricants.
Current edition approved Aug. 15, 1993. Published October 1993. Originally
published as D 3233 – 86. Last previous edition D 3233 – 92.
2 Annual Book of ASTM Standards, Vol 02.01.
3 Annual Book of ASTM Standards, Vol 05.01.
1
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
NOTICE: This standard has either been superceded and replaced by a new version or discontinued.
Contact ASTM International (www.astm.org) for the latest information.
6. Apparatus
6.1 Falex Pin and Vee Block Test Machine,4 illustrated in
Fig. 1, Fig. 2, and Fig. 3, fitted with 4500-lbf (20 000-N) gage
or 3000-lbf (13 350-N) gage.
6.2 Required for Calibration:
6.2.1 Allen Screw, with attached 10-mm Brinnell ball.5
6.2.2 Back-Up Plug.5
6.2.3 Standard Test Coupon,5 soft, annealed copper, Hb
37–39.
6.2.4 Brinnell Microscope, or equivalent.
6.2.5 Timer, graduated in seconds and minutes.
6.2.6 Rule, steel, 6-in. (approximately 150-mm) long.
7. Reagents and Materials
7.1 Standard Coined-Blocks,5 96 6 1° angle, AISI C-1137
steel, HRC 20 to 24, surface finish 5 to 10 µin. (1.3 3 10−7 to
2.5 3 10−7m), rms.
7.2 Standard Test Journals,5 1⁄4 in. (6.35 mm) outside
diameter by 11⁄4 in. (31.75 mm) long, AISI 3135 steel, HRB 87
to 91 on a ground flat surface, surface finish 5 to 10 µin.
(1.3 3 10−7 to 2.5 3 10−7 m) rms.
7.3 Locking Pins,5 1⁄2H brass, conforming to Specification
B 16.
7.4 Solvent, safe, nonfilming, nonchlorinated.
NOTE 4—Petroleum distillate and benzene, formerly used as solvents in
this method, have been eliminated due to possible toxic effects. Each user
should select a solvent that can meet applicable safety standards and still
thoroughly clean the parts.
8. Preparation of Apparatus
8.1 Cleaning:
8.1.1 Thoroughly clean the V-blocks, test journals, lubricant
cup, and supports for V-blocks and test journals by washing,
successively, with solvent selected in 7.4. Dry the V-blocks,
test journals, lubricant cup, and supports, by allowing the final
solvent to evaporate in air.
8.1.2 After cleaning, handle the test pieces with care to
prevent contamination. Particularly, avoid contact of fingers
with mating surfaces of V-blocks and test journals.
8.2 Assembly:
8.2.1 Insert the test journal into the test shaft and secure
with a new brass locking pin, as shown in Fig. 1 and Fig. 3.
8.2.2 Insert the V-blocks into the recesses of the loading
device and swing the V-blocks inward to contact the journal so
4 The Falex Pin and Vee Block Test Machine, available from the Falex Corp.,
1020 Airpark Dr., Sugar Grove, IL 60554 has been found satisfactory for this
purpose. A new model of this machine has been available since 1983. Certain
operating procedures are different for this new model. Consult instruction manual of
machine for this information.
5 Available from Falex Corp., 1020 Airpark Dr., Sugar Grove, IL 60554.
TABLE 1 Results of Cooperative Tests on Reference Fluids L-XI-1-2-A, B, C, D, EA TEST METHOD A
Labora-
tory Test
L-XI-1-2-A
Fail Load, lbf
L-XI-1-2-B
Fail Load, lbf
L-XI-1-2-C
Fail Load, lbf
L-XI-1-2-D
Fail Load, lbf
L-XI-1-2-E
Fail Load, lbf
Gage True Gage True Gage True Gage True Gage True
A 1
2
1200
1275
840
920
1200
1275
840
920
4500 +
4500 +
4100 +
4100 +
4300
4500 +
3950
4100 +
2600
2400
2100
1925
B 1
2
800
850
860
900
1000
950
1050
1025
4500 +
4500 +
4250 +
4250 +
4100
4300
3900
4100
2050
1950
2050
1950
C 1
2
725
650
990
910
775
750
1020
980
4500 +
4500 +
3200 +
3200 +
3950
4100
2900
3000
1350
1300
1460
1430
D 1
2
1400
1400
1050
1050
1100
1250
770
900
4500 +
4500 +
3500 +
3500 +
4500 +
4500 +
3500 +
3500 +
2900
2650
2150
1975
E 1
2
825
750
900
820
1000
925
1060
1000
4450
4450
4500 +
4500 +
4100
4150
4475
4500
1825
1825
1970
1970
F 1
2
1000B
990B
920B
910B
1000
1050
800
850
4500 +
4500 +
4500 +
4500 +
3500
2900
4500
3510
1850
1720
1900
1720
G 1
2
800
700
900
800
690
660
800
750
4000
3750
4275
4000
3325
3150
3625
3450
1430
1500
1600
1675
H 1
2
700
700
700
700
1000
1000
1000
1000
4500 +
4500 +
4500 +
4500 +
3750
4000
3750
4000
1900
1650
1900
1650
I 1
2
750
750
600
600
1250
1000
1000
800
4500 +
4500 +
3750 +
3750 +
4500 +
4500 +
3750 +
3750 +
1750
1750
1450
1450
Min Avg
Max Avg
Grand Avg
600
1050
854
775
1037
920
2950C
4488C
3809C
1445
2063
1796
Repeatability Reproducibility
s = 0.0624 S = 0.140
r = 0.179 (TL)D R = 0.402 (TL)D
AReference fluids used and described in Test Method D 2783.
BCalibration curves shifted.
CSix laboratories.
DTL = average true load, lbf, of sample tested.
D 3233
2
NOTICE: This standard has either been superceded and replaced by a new version or discontinued.
Contact ASTM International (www.astm.org) for the latest information.
that the V-grooves are aligned with the journal major axis, as
shown in Fig. 1 and Fig. 3.
8.2.3 Place 60 mL of test lubricant in the lubricant cup and
raise the cup so that the V-blocks are immersed in the test
lubricant. With highly viscous fluids open the jaws slightly to
ensure that the wear surfaces are covered with the lubricant.
TABLE 2 Results of Cooperative Tests on Reference Fluids L-XI-1-2-A, B, C, D, EA TEST METHOD B
Labora-
tory Test
L-XI-1-2-A
Fail Load, lbf
L-XI-1-2-B
Fail Load, lbf
L-XI-1-2-C
Fail Load, lbf
L-XI-1-2-D
Fail Load, lbf
L-XI-1-2-E
Fail Load, lbf
Gage True Gage True Gage True Gage True Gage True
A 1
2
1100
1100
750
750
1400
1400
1000
1000
4150
4350
3750
4000
4350
4150
4000
3750
2750
2200
2250
1750
B 1
2
670
670
750
750
940
670
1000
750
4200
3900
4000
3750
3900
4200
3750
4000
2000
1750
2000
1750
C 1
2
520
520
750
750
520
790
750
1000
4100 +
4100 +
3000 +
3000 +
4100
4100 +
3000
3000 +
1750
1750
1750
1750
D 1
2
1600
1600
1250
1250
1080
1080
750
750
4500 +
4500 +
3500 +
3500 +
4500 +
4500 +
3500 +
3500 +
3000
3300
2250
2500
E 1
2
700
700
750
750
925
925
1000
1000
3850
4150
4250
4500
3850
3650
4250
4000
1380
1850
1500
2000
F 1
2
1075B
1075B
1000B
1000B
950
950
750
750
3350
3500
4250
4500
3350
3050
4250
3750
1925
1560
2000
1500
G 1
2
660
660
750
750
660
800
750
1000
3500
3200
3750
3500
3000
2800
3250
3000
1550
1350
1750
1500
H 1
2
750
750
750
750
1000
1000
1000
1000
3500
4000
3500
4000
4250
4000
4250
4000
1500
1750
1500
1750
I 1
2
930
930
750
750
910
910
750
750
4400
4400
3750
3750
4400 +
4400 +
3750 +
3750 +
1800
1800
1500
1500
Min Avg
Max Avg
Grand Avg
750
1250
833
750
1000
875
3625C
4375C
3932C
3125D
4125D
3837D
1500
2375
1846
Repeatability Reproducibility
s = 0.0624 S = 0.137
r = 0.179 (TL)E R = 0.391 (TL)E
AReference fluids used and described in Test Method D 2783.
BCalibration curves shifted.
CSeven laboratories.
DSix laboratories.
ETL = average true load, lbf, of sample tested.
FIG. 1 Schematic Diagram of Falex Standard Pin and Vee Block Test Machine
D 3233
3
NOTICE: This standard has either been superceded and replaced by a new version or discontinued.
Contact ASTM International (www.astm.org) for the latest information.
8.2.4 Place the automatic loading device, with attached
gage, on the jaw arms.
9. Preparation of True Load Calibration Curve
9.1 On log-log paper (K & E 467080 or equivalent) draw a
straight-line plot of load, pounds-force (newtons) (ordinate),
versus indentation diameter, millimetres (abscissa), using the
data points shown below. Label this curve “True Load” (Note
5).
Load, lbf (N) Diameter, mm
(Ordinate) (Abscissa)
500 (2224) 2.62
1000 (4450) 3.42
1500 (6672) 4.00
2000 (8896) 4.47
NOTE 5—Fig. 3 shows the true-load calibration curve for the prescribed
4500-lbf (20 000-N) gage, prepared as covered in 9.1. Copies of Fig. 4, 8
by 11 in., are available at a nominal cost from ASTM. Although not
originally used in development of these test methods, the 3000-lb direct
reading load gage should be satisfactory providing results are corrected
and reported with respect to the true load (4500-lbf) reference line. Refer
to Test Method D 2670 for calibration of 3000-lb load gage.
10. Calibration of Load Gage 4500 lbf (20 000 N)
10.1 Remove the Allen set screw and 1⁄2-in. (12.70-mm) ball
from the left jaw socket (Fig. 5).
10.2 Insert the special Allen screw with the attached 10-mm
Brinnell ball into the working face of the left jaw. Adjust so
that the ball projects about 5⁄32 in. (approximately 4 mm) from
the face of the jaw.
10.3 Insert the back-up plug in the counterbore of the
right-hand jaw. Adjust so that the plug projects about 1⁄32 in.
(approximately 0.8 mm) from the face.
10.4 Support the standard test coupon so that the upper edge
of the coupon is about 3⁄32in. (approximately 2.5 mm) below
the upper surface of the jaws. Place a steel rule across the face
of the jaws. Adjust the Allen screw with the attached 10-mm
ball until the face of the jaws are parallel to the steel rule with
the test coupon in position for indentation.
10.5 With the test coupon in position for the first impres-
sion, place the load gage assembly on the level arms. Remove
the slack from the assembly by moving the ratchet wheel by
hand.
10.6 Place the loading lever on the ratchet wheel and actuate
the motor. Allow the motor to run until the load gage indicates
a load of 500 lbf (2224 N). A slight take-up on the ratchet
wheel is required to hold the load due to the ball sinking into
the test coupon. After a 500-lbf (2224-N) load is obtained, hold
for 1 min for the indentation to form.
10.7 Turn off the machine and back off the load until the test
coupon is free of the jaws. Advance the test coupon approxi-
mately 3⁄8 in. (approximately 9.5 mm). Additional indentations
should be separated by a minimum distance of 2.5 times the
diameter of the initial indentation. Check the alignment of the
jaws, and repeat the procedure described in 10.6 at gage loads
of 1000, 1500, and 2000 lbf (4448, 6672, and 8896 N).
10.8 Remove the load gage assembly and test coupon and
measure the diameter of each indentation to 0.01 mm with a
microscope. Make three measurements of the indentation
diameter, rotating the test coupon to ensure that no two
measurements represent the same points. Average the three
measurements of each impression and record.
10.9 Plot the four impression readings on the same log-log
plot of true load prepared as prescribed in 9.1 and shown as
Fig. 4. Draw a straight line through the four impression
FIG. 2 Falex Digital Pin and Vee Block Test Machine
FIG. 3 Exploded View of V-Blocks and Journal Arrangement,
Falex Pin and Vee Block Test Machines
D 3233
4
NOTICE: This standard has either been superceded and replaced by a new version or discontinued.
Contact ASTM International (www.astm.org) for the latest information.
readings and label the line “Actual Gage Load.”
NOTE 6—Currently, load gages are calibrated at the factory such that
the actual 4500-lb gage load is equivalent to true load. Periodic calibra-
tions should be made to ensure correct values are being reported for true
load.
TEST METHOD A
11. Determination of Actual Gage Load for Run-In
11.1 The procedure, Section 12, requires a run-in at an
actual gage load equivalent to 300-lbf (1334-N) true load
(264-lbf direct load). This actual gage load is obtained as
follows from the plot of actual gage load and true load prepared
in Sections 9 and 10: Locate 300 lbf (1334 N) on the true load
curve (264-lbf direct load). Through this point draw a vertical
line to intersect the actual gage load curve. Through this point
of intersection draw a horizontal line to the left-hand or
right-hand load scale and read the actual gage load value.
Record this actual gage load for run-in on a suitable reporting
form.
NOTE 7—A suitable reporting form for Test Methods A and B, and data
obtained on one of the reference fluids by one of the cooperating
laboratories, is shown in Table 3. Fig. 6 shows the calibration curves used
by the laboratory reporting the data in Table 3.
12. Procedure
12.1 Run-In:
12.1.1 Turn on “Heat Control” switch and heat test lubricant
to 120 6 5°F (51.7 6 3°C); then turn off the switch.
12.1.2 Remove slack from assembly by moving the ratchet
wheel by hand. At this setting the torque gage should read zero,
FIG. 4 Calibration Curve for 4500-lb Gage, Using 37-39 HB Copper Coupon
FIG. 5 Schematic Drawing of Calibration Accessories for
Falex Pin and Vee Block Test Machines
D 3233
5
NOTICE: This standard has either been superceded and replaced by a new version or discontinued.
Contact ASTM International (www.astm.org) for the latest information.
or be adjusted to zero.
12.1.3 Actuate the motor, engage the automatic loading
ratchet, and increase the load to a gage load equivalent to
300-lbf (1334-N) true load (264-lbf direct load), as determined
in 8.1. Disengage the loading ratchet, start the timer, and allow
the machine to run at this loading for a 5-min run-in period.
NOTE 8—Maintain load at near constant by taking up the load manually
or automatically by means of the ratchet wheel if necessary.
12.2 Test:
12.2.1 Re-engage the automatic loading ratchet and leave it
engaged until failure (Note 9) or until the highest indicated
actual gage reading is reached. Stop the motor at failure or at
the highest indicated actual gage load when no failure is
obtained. Record the gage load at failure. Record 4500 lbf
(20 000 N) if no failure is obtained.
NOTE 9—Failure is indicated by (a) breakage of shear pin or test pin, or
(b) inability to take up the load automatically by means of the ratchet
wheel.
12.2.2 Using the calibration curves prepared in Sections 9
and 10, determine and record the true load failure equivalent to
the actual gage load failure, or, if no failure, the true load
equivalent to the highest indicated actual gage load, with a plus
( + ) sign after the true load value.
NOTE 10—To convert actual gage load to true load, locate the gage load
on the actual gage load curve. Through this point draw a vertical line to
intersect the true load curve. Through this point of intersection draw a
horizontal line to the left-hand or right-hand load scale and read the true
load value.
TEST METHOD B
13. Determination of Actual Gage Load for Run-In and
250-lbf (1112-N) True Load Increments
13.1 The procedure, Section 14, requires a run-in at an
actual gage load equivalent to 300-lbf (1334-N) true load
(264-lbf direct load), and testing at incremental gage loadings
equivalent to 250-lbf (1112-N) true load (224-lbf direct load)
over the range from 500 to 4500-lbf (2224 to 20 000-N) true
load (412 to 2885-lbf direct load). Determine the equivalent
actual gage loads as prescribed in Section 8, and Test Method
A. Record on a suitable reporting form, such as shown in Table
3.
14. Procedure
14.1 Run-In—Use same procedure as prescribed in 12.1,
Test Method A.
14.2 Test:
14.2.1 Re-engage the automatic loading ratchet and leave it
engaged until the actual gage reading is equivalent to 500-lbf
TABLE 3 Suggested Report Form, Test Methods A and B, Showing Data
Operator: Laboratory A
Test Sample: L-XI-1-2-E Gage Type:
Calibration Indentation Results: 500 lbf = 2.13 mm
1000 lbf = 2.90 mm
1500 lbf = 3.53 mm
2000 lbf = 4.06 mm
TEST METHOD A
Run-In: 5 min at 520 lbf actual gage load, equivalent to 300 lbf true load.
Actual Gate Load,
lbf, at Failure
Equivalent True Load,
lbf at Failure
Test No. 1 2600 2100
Test No. 2 2400 1925
TEST METHOD B
Run-In: 5 min at 520 lbf actual gage load, equivalent to 300 lbf true load.
True Load, lbf Equivalent ActualGage Load, lbf Test No. 1 Test No. 2
500 800 pass pass
750
本文档为【ASTM D3233 – 93 R98】,请使用软件OFFICE或WPS软件打开。作品中的文字与图均可以修改和编辑,
图片更改请在作品中右键图片并更换,文字修改请直接点击文字进行修改,也可以新增和删除文档中的内容。
该文档来自用户分享,如有侵权行为请发邮件ishare@vip.sina.com联系网站客服,我们会及时删除。
[版权声明] 本站所有资料为用户分享产生,若发现您的权利被侵害,请联系客服邮件isharekefu@iask.cn,我们尽快处理。
本作品所展示的图片、画像、字体、音乐的版权可能需版权方额外授权,请谨慎使用。
网站提供的党政主题相关内容(国旗、国徽、党徽..)目的在于配合国家政策宣传,仅限个人学习分享使用,禁止用于任何广告和商用目的。