Sepax Technologies, Inc.
Delaware Technology Park
5 Innovation Way, Newark, DE 19711, USA
Phone: (302) 366-1101; Fax: (302) 366-1151
Toll Free: 1-877-SEPAX-US; www.sepax-tech.com
Sepax Mono-GPC Column Manual
Column Information
Mono GPC resins have been specifically designed for high
resolution gel permeation separation of synthetic polymers. These
novel packing materials are based on highly cross-linked
polystyrene / divinylbenzene (PS/DVB) particles with very
narrow particle size and pore size distributions. Their uniform
pore size distribution offers near linear calibration curves
covering wide molecular weight range. The highly cross-linked
porous particles provide excellent chemical and physical stability
and permit easy transfer across the full range of organic solvents
with little change in the shape of the calibration curve or the
efficiency of the columns. Their high pore volume enables the
high resolution of polymer separation. Mono GPC column
packings can be the resins with an individual pore size or a
mixture of a series of pore sizes.
Characteristics
Support: Spherical, PS/DVB particles
Pore size: 100, 300, 500 and 1,000 Å
Particle size: 5 and 10 µm
Pore volume: ~1.0 mL/g
Phase structure: hydrophobic
Chemical composition: polystyrene/divinylbenzene
Column Stability and Performance
Mono GPC columns are highly stable over variety of operation
conditions. They are stable to resist high temperature up to 145˚ C
in nearly all organic solvents. Solvents can be changed without
damaging the column. Mono GPC columns have long life time –
negligible deterioration after 3 month of standard usage. The
narrow distributed pore size enables linear calibrations that can
maximize precision of molecular weight calculations. Mon GPC
offers a wide range of porosities from 100Å to 1000Å as well as
mixed beds. Mono-dispersed particles achieve high efficient
separation and their high pore volume results in greater resolution.
A typical test chromatogram for quality control is shown in Figure
1 for a 4.6x250mm Mono GPC-300 column.
Safety Precaution
Mono GPC columns are normally operated under high
pressure. Loose connections will cause leaking of organic solvents
and injected samples, all of which should be considered as the
hazards. In the case of leaking, proper gloves should be worn for
handling the leaked columns. When open the columns, proper
protections should be used to avoid inhalation of the small
polymer particles.
A
U
0.00
0.04
0.08
0.12
0.16
0.20
0.24
0.28
Min
0 1 2 3 4 5 6 7 8 9 10
Column: Mono GPC-300 (5 µm, 300 Å)
4.6x300 mm
Mobile phase: Tetrahydrofuran
Flow rate: 1.0 mL/min
Detection (UV): 254 nm
Injection Volume: 5 µL
Peptides: Toluene (5% in THF)
A
U
0.00
0.04
0.08
0.12
0.16
0.20
0.24
0.28
Min
0 1 2 3 4 5 6 7 8 9 10
A
U
0.00
0.04
0.08
0.12
0.16
0.20
0.24
0.28
Min
0 1 2 3 4 5 6 7 8 9 10
Column: Mono GPC-300 (5 µm, 300 Å)
4.6x300 mm
Mobile phase: Tetrahydrofuran
Flow rate: 1.0 mL/min
Detection (UV): 254 nm
Injection Volume: 5 µL
Peptides: Toluene (5% in THF)
Figure 1. Toluene is used to test the column for quality control.
Column Installation and Operation
When column is shipped or not in use, it is always capped at
both ends. When install the column to the system, first remove the
end caps. Make the flow direction as marked on the column.
Unless a user has special purpose to reverse the flow direction, for
example, removal of the inlet pluggage, follow the flow direction
as labeled. Column connections are an integral part of the
chromatographic process. If ferrules are over tightened, not set
properly, or are not specific for the fitting, leakage can occur. Set
the ferrules for column installation to the HPLC system as
follows:
(a) Place the male nut and ferrule, in order, onto a 1/16" o.d.
piece of tubing. Be certain that the wider end of the ferrule is
against the nut.
2
(b) Press tubing firmly into the column end fitting. Slide the nut
and ferrule forward, engage the threads, and fingertighten the nut.
(c) While continuing to press the tube firmly into the endfitting,
use a 1/4" wrench to tighten the nut 90 degrees past
fingertightness.
(d) Repeat this coupling procedure for the other end of the
column.
New Mono GPC columns are shipped in tetrahydrofuran
(THF). During stocking and shipping, the polymer packing could
be dried out. It is recommended that 10-20 column volumes of
pure organic solvents, such as THF be purged to activate the
column. Flush the column with your mobile phase with gradual
increasing the flow rate from 0.1 mL/min to your operation
condition, until the baseline is stable. If the column backpressure
and baseline fluctuate, this might be due to the air bubbles trapped
inside the column. Flush the column with higher flow rate for 2-5
minutes, for example 1.0 mL/min for 4.6x300mm.
Samples and Mobile Phases
To avoid clogging the column, all samples and solvents
including buffers should be filtered through 0.45 µm or 0.2 µm
filters before use. They are stable to resist high temperature up to
145˚ C in nearly all organic solvents. Typical solvent systems
include tetrahydrofuran (THF), chloroform, dimethylacetamide
(DMAC), dimethylformamide (DMF), trichlorobenzene (TCB),
N-methylpyrrolidone (NMP), Hexafluoroisopropanol (HFIP) and
Toluene. Solvents can be changed without damaging the column.
Before changing solvents please confirm that your column is
compatible with your new mobile phase. See our solvent
compatibility chart for help or call us for technical assistance.
Always purge your column into a new solvent at .2ml/min until
two full column volumes have passed through the column.
Column Care
PH Wide pH range - mobile phase & pH values
from 0 to 14. Avoid storing the column below pH 1 or above 12
when not use. The extreme pH would damage the stainless steel
column tube and frits for long time storage.
Pressure Even though Mono GPC columns can operate
at pressure up to 5,000 psi, the normal operation is usually under
3,000 psi. Continuous use at high pressure may eventually
damage the column as well as the pump. Since the pressure is
generated by the flow rate. The maximum flow rate is limited by
the backpressure. It is expected that the backpressure might
gradually increase with its service. A sudden increase in
backpressure suggests that the column inlet frit might be plugged.
In this case it is recommended that the column be flushed with
reverse flow in an appropriate solvent.
Temperature The maximum operating temperature is
145oC. Continuous use of the column at higher temperature
(>145oC) can damage the column, especially under pure organic
solvents.
Storage When not in use for extended time, it is
recommended to store the column in pure THF solvent.
Avoiding Tailing and/or Adsorption Based on
styrene/divinylbenzene, Mono GPC resins have a large number of
aromatic rings inherent in the packing’s structure will give unique
responses to certain types of samples that contain aromatic rings
or atoms such as O or N with unshared electron pairs. Those
samples have the potential to be strongly retained and/or tail on
the Mono GPC columns unless there is a competing electron-rich
solvent in the mobile phase. Thus, to obtain sharper peaks with
less tailing and good resolution, you can “adjust” the surface
chemistry with a competing electron-rich solvent like acetonitrile
or use a mobile phase additive such as triethylamine (TEA) or n-
butylamine which can coordinate with the aromatic rings of the
packing material creating a less electron-dense surface chemistry.
For certain separations it is also possible to use sodium acetate to
modify peak shape and retention intensity. In like manner, using
low percentages of glycerol, 2-propanol, or other similarly
hydrophilic hydroxylated solvents reduces the net effective
surface hydrophobicity. It is recommended to use quantities of
0.5-2.0% of TEA or ethylene glycol, or 0.01M Na Acetate, and
anywhere from 2.0-100% of solvents such as CH3CN, CH3OH, or
2-propanol.
Frit Fouling If you notice a calibration change after
significant use, you may need a clean frit particularly on the
column inlet. If it the original inlet frit clogs, it will contribute to
shearing of high MW polymer.
Mono GPC Products
5 µm Phases 300x4.6 300x7.8
Mono GPC-100 (100 Å) 230100-4630 230100-7830
Mono GPC-300 (300 Å) 230300-4630 230300-7830
Mono GPC-500 (500 Å) 230500-4630 230500-7830
Mono GPC-1000
(1000 Å)
231000-4630 231000-7830
Mono GPC-5MP
(Mixed Bed 100-1000 Å)
231135-4630 231135-7830
10 µm Phases 300x4.6 300x7.8
Mono GPC-100 (100 Å) 240100-4630 240100-7830
Mono GPC-300 (300 Å) 240300-4630 240300-7830
Mono GPC-500 (500 Å) 240500-4630 240500-7830
Mono GPC-1000
(1000 Å)
241000-4630 241000-7830
Mono GPC-10MP
(Mixed Bed 100-1000 Å)
241135-4630 241135-7830
Sepax Mono-GPC Column Manual
Column Information
Characteristics
Column Stability and Performance
Safety Precaution
Column Installation and Operation
Samples and Mobile Phases
Column Care
Mono GPC Products
本文档为【Sepax Mono-GPC Column Manual】,请使用软件OFFICE或WPS软件打开。作品中的文字与图均可以修改和编辑,
图片更改请在作品中右键图片并更换,文字修改请直接点击文字进行修改,也可以新增和删除文档中的内容。
该文档来自用户分享,如有侵权行为请发邮件ishare@vip.sina.com联系网站客服,我们会及时删除。
[版权声明] 本站所有资料为用户分享产生,若发现您的权利被侵害,请联系客服邮件isharekefu@iask.cn,我们尽快处理。
本作品所展示的图片、画像、字体、音乐的版权可能需版权方额外授权,请谨慎使用。
网站提供的党政主题相关内容(国旗、国徽、党徽..)目的在于配合国家政策宣传,仅限个人学习分享使用,禁止用于任何广告和商用目的。